Product outline
Specifications
Analog section|Display section|Setting section|External signal|General performance|Attachments
Analog section | |
Sensor excitation voltage | 10/2.5V DC ±10%; Output current: within 30mA |
---|---|
Signal input range | -3.0 to +3.0mV/V |
Accuracy |
Non-linearity: within 0.02%/FS (at 3mV/V input) Zero drift: within 0.5μV/°C RTI Gain drift: within 25ppm/°C |
A/D converter |
Speed: 100 times/sec. Resolution: 16 bit (binary) |
Analog filter |
Selectable from 4, 10, 100, 3k Hz Low-pass filter (-6dB/oct.) |
Peak hold function (High-speed analog hold method) |
At 3mV/V input: Performance response speed: approx. 1 kHz (at sine wave input, analog filter 3 kHz) Accuracy: 0.1% FS or less, Reset time: 50μs or less |
Voltage output | Analog voltage proportional to sensor input; Output level: approx. 2V per 1mV/V input |
SI/F output | Output data: Indicated value, status |
Display section | |
Display unit |
15mm character height; 7-segment red LED numerical display (5 digits) Digit numerals: ±8.8.8.8.8 Indicated value: -19999 to +19999 Decimal points: selectable decimal point displaying position |
---|---|
Display items |
Status display: 5 red LED display for HI, OK, LOW, PEAK, HOLD Display times/sec: selectable from 3, 6, 13 and 25 times/sec. |
Setting section | |
Setting items |
Calibration: Zero/ Span calibration (actual load/equiv. input calibration) HI limit, LO limit, HI/LO comparison mode, hysteresis, digital offset, near zero, digital filter, analog filter, motion detect, zero tracking, hold mode, auto print, hold value print, LOCK, min scale, display frequency, excitation voltage, BCD data update rate, RS-232C, D/A converter zero setting, D/A converter full scale setting |
---|
External signal | |
External I/O signal | HI limit relay (N/O) output, LO limit relay (N/O) output, digital zero input, hold input |
---|---|
Interface |
SIF: 2-wire type serial interface 232: RS-232C communication interface (Optional) BCO: BCD parallel data output interface (Optional) DAV: D/A converter interface (voltage output) (Optional) DAI: D/A converter interface (current output) (Optional) * Only 1 option can be installed. |
General performance | |
Power supply voltage | 100V to 240V AC (+10% -15%) (free power supply), 50/60 Hz, 12 to 24V DC (±15%) (Please specify when ordering.) |
---|---|
Power consumption | 15W max |
Operating conditions |
Temperature: Working temperature range -10 to +40°C; Storage temperature range -40 to +80°C Humidity: 85% RH or less (non-condensing) |
Dimensions | 96(W)×96(H)×135(D) mm (protruding parts not included) |
Panel cutout dimensions | 92×92(+1-0)mm |
Weight | Approx. 0.9kg |
CE marking certification |
EMC directives: EN61326-1 Safety standard: EN61010-1, EN62311 |
Attachments |
AC input cord (nominal rating 125A) 3m * ×1 3P-2P conversion plug for AC input cord * ×1 Mini-driver ×1 Ferrite core for sensor cable ×1 Ferrite core for AC input cord * ×1 BCD output connector (when BCD output option is installed) ×1 Operation manual ×1 * Included only for AC power supply specification |
---|
Option
Model | |
BCO | BCD parallel data output interface |
---|---|
232 | RS-232C communication interface |
DAV | D/A converter interface (voltage output) |
DAI | D/A converter interface (current output) |
Optional accessories
Model | |
CAAC3P-B3 | AC Supply cord 3m |
---|---|
CAAC3P-CEE7/7-B2 | AC Supply cord (Voltage resistance: 250V) 2m |
CN3P-2P | 3P-2P converter plug for AC input cord |
CN35 | D-Sub25p connector for RS-232C |
CN21 | BCD output connector |
TSU01 | Lightning surge unit |
TSU03 | Lightning surge unit, DC specification |
Download
Product catalogue(PDF)
|
GC2017_F340A(239 KB) |
---|
咨询 - F340A (Force Indicator)