Share:


USLC (Compact Compression Type)



* Supported only for USLC-100KN

Product outline

  • Can measure up to 50 kN with diameter of 22 mm, and 100 kN with diameter of 38 mm!

  • Accuracy 0.3%; astounding among those compact & high-capacity load cells

  • Easy & fast replacement from competitors’ load cells

     

Specifications

Specification USLC-50KN USLC-100KN
Rated capacity 50 kN 100 kN
Rated output Approx. 2 mV/V —
Safe overload 150% R.C. —
Zero balance ±10% R.O. —
Non-linearity 0.3% or less R.O. —
Hysteresis 0.3% or less R.O. —
Repeatability 0.1% or less R.O. —
Compensated temperature range -10 to +60℃ —
Safe temperature range -20 to +70℃ —
Temperature effect on zero 0.5% R.O./10℃ or below —
Temperature effect on span 0.5% R.O./10℃ or below —
Input resistance Approx. 700 Ω —
Output resistance Approx. 700 Ω —
Recommended excitation voltage 5 V —
Maximum excitation voltage 10 V —
Insulation resistance (DC 50 V) 1000 MΩ or more —
Cable φ2 4-conductor flexible cable (3 m); Cable end: 5 wires φ3.8 4-conductor flexible cable (3 m); Cable end: 5 wires
Load cell material Stainless steel —
Deflection at rated 0.08 mm 0.07 mm
Natural frequency 39 kHz 16.3 kHz
Weight 40 g 210 g
RoHS directives 2011/65/EU (EU)2015/863 —

* Due to the characteristics of the stainless steel being used for, it is possible that it gets rusty on its surface.There is functionally no problem.


 

Application example, Wiring diagram

External dimension

 



 Inquiry - USLC (Compact Compression Type)